beautypg.com

Deluxe version ift displays and controls, Deluxe version ift displays and controls -2, World class 3000 – Emerson Process Management 3000 User Manual

Page 72: Instruction manual

background image

Instruction Manual

IB-106-300NH Rev. 4.2
July 2002

5-2 General User Interface (GUI) Operation

Rosemount Analytical Inc. A Division of Emerson Process Management

World Class 3000

5-2

DELUXE VERSION IFT DISPLAYS AND CONTROLS

EN

TE

R

ESC

DA

TA

CA

L

SET

UP

HE

LP

1

2

3

4

5

6

7

8

9

INTERNAL VIEW

CAL
TGH
TGL

10

11

12

13

21190003

EXTERNAL VIEW

IFT COVER DOOR
SHOWN FOR
REFERENCE.

NOTE:

Figure 5-1. Deluxe Version IFT Displays and Controls

Figure 5-1

Index No.

Control/LED

Description

1

LCD Display

Top line displays system status, menu, and probe number.

2

HELP

Context sensitive HELP is displayed when this key is pressed.

3

DATA

DATA key is used to access DATA menu.

4

CAL

CAL key used to access CALIBRATE menu.

5

SETUP

SETUP key used to access SETUP menu.

6

ESC

The escape key is used to exit to a high level menu or to abort a
parameter change.

7

The decrease key is used to move the cursor (asterisk) when scrolling
through lists or to decrease a parameter value.

8

The increase key is used to move the cursor (asterisk) when scrolling
through lists or to increase a parameter value.

9

ENTER

The ENTER key is used to select a lower level menu, initiate calibra-
tion, or select a parameter to change.

10

LED Display

Indicates current O

2

or calibration gas value.

11

CAL

Calibration in progress indicator light.

12

TGH

High calibration gas indicator light. High calibration gas is being used in
calibration process.

13

TGL

Low calibration gas indicator light. Low calibration gas is being used in
calibration process.