Figure 2-8, To figure 2-8, World class 3000 – Emerson Process Management 3000 User Manual

Page 45: Instruction manual, Iv current/voltage selector switch

background image

Instruction Manual

IB-106-300NH Rev. 4.2

July 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Installation 2-13

World Class 3000

BK

GN

WH

PU

SHIELD

SHIELD

CAL RET
NO GAS

LO GAS

HI GAS
IN GAS

SHIELD

SHIELD

STACK TC -
STACK TC +

PROBE MV

+

PROBE MV -

PROBE TC +

PROBE TC -

RD

YE

BL

OR

BL

OR

YE
RD

INTELLIGENT FIELD

TRANSMITTER IFT 3000

WORLD CLASS

PROBE

1 2 3 4 5 6 7 8

BL

PROBE

MV

-

PROBE

MV

+

PROBE

TC

+

PROBE

TC

-

OR

YE

RD

GN
E

R

H

WH

BK

GN

CELL

-VE

OR

CELL

+VE

YE

CHROMEL

RD

ALUMEL

GN

BK

BK

}

HEA

TER

PROBE JUNCTION

BOX WIRING

STACK TC WIRING AS REQUIRED.

SPECIAL PROBE CABLE BETWEEN PROBE

AND IFT BY ROSEMOUNT.

INSTALL JM1 ON INTERCONNECT BOARD.

INSTALL JM6 ON MICROPROCESSOR

BOARD.

IF STACK TEMPERATURE NOT USED.

IF MPS 3000 NOT USED.

1 RELAY PER PROBE AVAILABLE FOR

CALIBRATION STATUS INDICATION. (48 V

max, 100 mA max)

CURRENT/VOLTAGE SELECTOR SWITCH

MUST BE SELECTED TO CURRENT (I) FOR

HART COMMUNICATIONS APPLICATIONS.

JUMPER JM7 INFORMATION APPEARS IN

FIGURE 2-9.

NOTES:

ALWAYS DISCONNECT LINE VOLTAGE

FROM INTELLIGENT FIELD TRANSMITTER

BEFORE CHANGING JUMPERS.

LINE

VOLTAGE

SECTION

JUMPER

(INSTALL)

PROBE HEATER

VOLTAGE SECTION

JUMPER

(INSTALL)

120 V.A.C.

100 V.A.C.

220 V.A.C.
240 V.A.C.

J1

J5

J6

J1

JM1

JM7

JM6

J2

J3

J4

J5

J6

J7

J8

J9

3D39122G REV

POWER SUPPLY BOARD

3D39513G

MICROPROCESSOR

BOARD

3D39120G REV

INTERCONNECT BOARD

JM8, JM7, JM1

JM3, JM7, JM2

JM6, JM5, JM2
JM6, JM5, JM1

WORLD CLASS PROBE

WORLD CLASS "DIRECT

REPLACEMENT" PROBE

218 PROBE

JM10

JM9

JM9

JUMPER CONFIGURATION

5 CONDUCTOR SHIELDED CABLE

PER PROBE #16 AWG BY CUSTOMER

E

N

L

LINE

VOLTAGE

MPS TERMINATION BOARD

MPS 3000 MULTIPROBE CALIBRATION GAS SEQUENCER (OPTIONAL)

E

R

H

L
E

N

LINE

VOLTAGE

J13 J14

J15

J16

J17 J18

J12

CAL

RET

HI

GAS

IN

CAL

NO

GAS

CAL

RET

HI

GAS

IN

CAL

NO

GAS

CAL

RET

HI

GAS

IN

CAL

NO

GAS

CAL

RET

HI

GAS

IN

CAL

NO

GAS

LOW

GAS

LOW

GAS

LOW

GAS

LOW

GAS

NC C NO NC C NO NC C NO NC C NO

L

N

L

N

LINE OUT

LINE IN

J10

J11

PROBE 1

PROBE 2

PROBE 3

PROBE 4

PROBE 1

PROBE 2

PROBE 3

PROBE 4

PROBE

1

SOLENOID

PROBE

2

SOLENOID

PROBE

3

SOLENOID

PROBE

4

SOLENOID

HIGH

GAS

SOLENOID

LOW

GAS

SOLENOID

PRESSURE SWITCH

I

V

CURRENT/VOLTAGE

SELECTOR SWITCH

27270011

Figure 2-8. Wiring Layout for IFT Systems without HPS