Multiprobe calibration gas sequencer installation, Figure 2-18. mps module -22, World class 3000 – Emerson Process Management 3000 User Manual
Page 54: Instruction manual, Figure 2-18. mps module
Instruction Manual
IB-106-300NH Rev. 4.2
July 2002
2-22 Installation
Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
2-4
MULTIPROBE CALIBRATION GAS
SEQUENCER INSTALLATION
a. Mechanical
Installation
The outline drawing of the MPS module in
Figure 2-18 shows mounting centers and
clearances. The box is designed to be
mounted on a wall or bulkhead. The MPS
module should be installed no further than
300 feet (91 mz) piping distance from the
probe, and no more than 1000 feet (303 m)
cabling distance from the IFT. Install the
MPS module in a location where the ambi-
ent temperature is between -20° and 160°F
(-30° and 71°C).
b. Gas
Connections
Figure 2-19 shows the bottom of the MPS
where the gas connections are made. 1/4
inch threaded fittings are used.
1. Connect the reference air supply to
INSTR. AIR IN. The air pressure regu-
lator valve is set at the factory to 20 psi
(138 kPa). If the reference air pressure
should need readjustment, turn the
knob on the top of the valve until the
desired pressure is obtained.
2. Connect the high O
2
calibration gas to
HIGH GAS. The calibration gas pres-
sure should be set at 20 psi (138 kPa).
HIGH CAL
GAS IN
LOW CAL
GAS IN
CAL GAS
OUT
REF AIR
OUT
INSTR
AIR
REF AIR
OUT
REF AIR
OUT
REF AIR
OUT
CAL GAS
OUT
CAL GAS
OUT
CAL GAS
OUT
PROBE 1
PROBE 2
PROBE 3
PROBE4
0.84 (21.34)
27270013
1.96 (49.78)
4.21 (106.93)
3.09 (78.49)
5.25 (133.35)
5.54 (140.72)
14.00 (355.60) REF
12.00
(304.80)
12.00
(304.80)
10.00
(254.00)
NOTE: DIMENSIONS ARE IN INCHES
WITH MILLIMETERS IN
PARENTHESES.
Figure 2-18. MPS Module