World class 3000 – Emerson Process Management 3000 User Manual
Page 111
Instruction Manual
Appendix A Rev. 3.8
July 2002
Rosemount Analytical Inc. A Division of Emerson Process Management
Appendices A-13
World Class 3000
Table A-2. Fault Finding (Continued)
Symptom
Check
Fault
Remedy
2. System responds to
oxygen concentration
changes but does not
give correct indication
Recorder or remote
indicator
Calibration error
Recalibrate recorder or
indicator. Reference Re-
corder Instruction Manual.
System calibration
Calibration error
Recalibrate system.
Probe mounting and con-
dition of duct
Air ingress into duct
Stop air leaks or resite
probe.
Good response, with
incorrect indication
Cell mV input to
electronics
Failure of electronics
Check electronics
package.
Calibration gas input port
Blocked port
Clean port. If the flue gas
is condensing in the cali-
bration gas line, insulate
the back of the probe.
Make sure that the cali-
bration gas line is capped
between calibrations, or a
check valve is installed.
3. System does not give
accurate indication of
applied calibration gas
Ceramic diffusion element Diffusion element cracked,
broken, or missing
Replace diffusion element.
4. System takes a long
time to return to flue
gas value after calibra-
tion gas is turned off
Diffusion element
Plugged diffusion element Change diffusion element
or snubber diffusion
element.
Check calibration gas fit-
ting.
Leak in gasket, or calibra-
tion gas fittings or tubing.
Repair leak.
Check calibration gas
hoses or tubing.
Leak in calibration gas
hose or tubing.
Repair hose or tubing
.
Cell flange corrugated
seal.
Leaking seal.
Replace seal.
5. Probe passes calibra-
tion, but appears to
read high.
If equipped with a
n abra-
sive shield, check flange
mounting gasket.
Leaking gasket.
Replace gasket.
6. Probe passes calibra-
tion, but appears to
read low.
Flow calibration gas to the
probe until reading stabi-
lizes. Shut off calibration
gas and note the time re-
quired to return to process
gas values. Time should
be <20 sec. (<40 sec. for
hazardous area probes).
Plugged diffuser element
from high process par-
ticulate may cause a cali-
bration shift.
Replace diffuser.
A