beautypg.com

World class 3000 – Emerson Process Management 3000 User Manual

Page 111

background image

Instruction Manual

Appendix A Rev. 3.8

July 2002

Rosemount Analytical Inc. A Division of Emerson Process Management

Appendices A-13

World Class 3000

Table A-2. Fault Finding (Continued)

Symptom

Check

Fault

Remedy

2. System responds to

oxygen concentration
changes but does not
give correct indication

Recorder or remote
indicator

Calibration error

Recalibrate recorder or
indicator. Reference Re-
corder Instruction Manual.

System calibration

Calibration error

Recalibrate system.

Probe mounting and con-
dition of duct

Air ingress into duct

Stop air leaks or resite
probe.

Good response, with
incorrect indication

Cell mV input to
electronics

Failure of electronics

Check electronics
package.

Calibration gas input port

Blocked port

Clean port. If the flue gas
is condensing in the cali-
bration gas line, insulate
the back of the probe.
Make sure that the cali-
bration gas line is capped
between calibrations, or a
check valve is installed.

3. System does not give

accurate indication of
applied calibration gas

Ceramic diffusion element Diffusion element cracked,

broken, or missing

Replace diffusion element.

4. System takes a long

time to return to flue
gas value after calibra-
tion gas is turned off

Diffusion element

Plugged diffusion element Change diffusion element

or snubber diffusion
element.

Check calibration gas fit-
ting.

Leak in gasket, or calibra-
tion gas fittings or tubing.

Repair leak.

Check calibration gas
hoses or tubing.

Leak in calibration gas
hose or tubing.

Repair hose or tubing

.

Cell flange corrugated
seal.

Leaking seal.

Replace seal.

5. Probe passes calibra-

tion, but appears to
read high.

If equipped with a

n abra-

sive shield, check flange
mounting gasket.

Leaking gasket.

Replace gasket.

6. Probe passes calibra-

tion, but appears to
read low.

Flow calibration gas to the
probe until reading stabi-
lizes. Shut off calibration
gas and note the time re-
quired to return to process
gas values. Time should
be <20 sec. (<40 sec. for
hazardous area probes).

Plugged diffuser element
from high process par-
ticulate may cause a cali-
bration shift.

Replace diffuser.

A