Appendices D-9

World Class 3000

LEGEND FOR FIGURE D-6

1. Enclosure

2. Screw

3. Plug

4. Cable

Grip

5. Fitting

6. Hose

Adapter

7. Pressure

Switch

8. Plug

9. Solenoid

Valve

10. Screw

11. Manifold

12. Washer

13. Screw

14. Gasket

15. Outer

Cover

16. Inner

Cover

17. Flowmeter,

10

SCFH

18. Flowmeter,

2.0

SCFH

19. Bracket

20. Screw

21. Hose

Adapter

22. 1/8 in. Hose

23. Nut

24. Lockwasher

25. Washer

26. Washer

27. Screw

28. Nut

29. Washer

30. Washer

31. Cover Stop Slide

32. Screw

33. Washer

34. Termination

Board

35. Standoff

36. Mounting

Bracket

37. Cover

Plate

38. Washer

39. Screw

40. Fuseholder

41. Plastic

Nut

42. Bushing

43. Pressure

Gauge

44. Bolt

45. Washer

46. Drain

Valve

47. 1/8 in. Impolene Tubing

48. Connector

49. Elbow

50. Pressure

Regulator

51. Hose

Adapter

52. 1/4 in. Tube

53. Screw

54. Washer

55. Inner

Enclosure

56. Washer

57. Screw

58. Power

Supply

59. Washer

60. Nut

61. Screw

62. Washer

63. Solenoid

64. Solenoid

e. Loosen the retaining ring in the middle of

the solenoid and remove the top part.

f. With a spanner wrench or padded pliers,

remove the remaining part of the solenoid
from the manifold (11).

g. Separate the new solenoid and screw the

smaller part into the manifold.

h. Place the top part of the solenoid into posi-

tion and tighten the retaining ring.

i.

Connect the plug to the proper receptacle
on the termination board (34).

j.

Close and secure the inner cover (16) with
two captive screws. Close and secure the
outer cover (15) with two captive screws.

D-9

PRESSURE REGULATOR MAINTENANCE

a. Pressure

Adjustments

Pressure regulator (50, Figure D-6) is fac-
tory set to 20 psi (138 kPa). Should the
pressure need to be changed or adjusted,
use the knob on top of the pressure
regulator.

b. Condensation

Drain

To drain excess moisture from the internal
gas circuit of the MPS, periodically loosen
drain valve (46) on the bottom of pressure
regulator (50). The moisture will flow
through vinyl tubing drain (47) on the bottom
of pressure regulator (50) and exit the bot-
tom of MPS enclosure (1).

D