Pressure regulator maintenance, World class 3000, Instruction manual – Emerson Process Management 3000 User Manual
Page 145
Instruction Manual
Appendix D Rev. 2.4
July 1998
Rosemount Analytical Inc. A Division of Emerson Process Management
Appendices D-9
World Class 3000
LEGEND FOR FIGURE D-6
1. Enclosure
2. Screw
3. Plug
4. Cable
Grip
5. Fitting
6. Hose
Adapter
7. Pressure
Switch
8. Plug
9. Solenoid
Valve
10. Screw
11. Manifold
12. Washer
13. Screw
14. Gasket
15. Outer
Cover
16. Inner
Cover
17. Flowmeter,
10
SCFH
18. Flowmeter,
2.0
SCFH
19. Bracket
20. Screw
21. Hose
Adapter
22. 1/8 in. Hose
23. Nut
24. Lockwasher
25. Washer
26. Washer
27. Screw
28. Nut
29. Washer
30. Washer
31. Cover Stop Slide
32. Screw
33. Washer
34. Termination
Board
35. Standoff
36. Mounting
Bracket
37. Cover
Plate
38. Washer
39. Screw
40. Fuseholder
41. Plastic
Nut
42. Bushing
43. Pressure
Gauge
44. Bolt
45. Washer
46. Drain
Valve
47. 1/8 in. Impolene Tubing
48. Connector
49. Elbow
50. Pressure
Regulator
51. Hose
Adapter
52. 1/4 in. Tube
53. Screw
54. Washer
55. Inner
Enclosure
56. Washer
57. Screw
58. Power
Supply
59. Washer
60. Nut
61. Screw
62. Washer
63. Solenoid
64. Solenoid
e. Loosen the retaining ring in the middle of
the solenoid and remove the top part.
f. With a spanner wrench or padded pliers,
remove the remaining part of the solenoid
from the manifold (11).
g. Separate the new solenoid and screw the
smaller part into the manifold.
h. Place the top part of the solenoid into posi-
tion and tighten the retaining ring.
i.
Connect the plug to the proper receptacle
on the termination board (34).
j.
Close and secure the inner cover (16) with
two captive screws. Close and secure the
outer cover (15) with two captive screws.
D-9
PRESSURE REGULATOR MAINTENANCE
a. Pressure
Adjustments
Pressure regulator (50, Figure D-6) is fac-
tory set to 20 psi (138 kPa). Should the
pressure need to be changed or adjusted,
use the knob on top of the pressure
regulator.
b. Condensation
Drain
To drain excess moisture from the internal
gas circuit of the MPS, periodically loosen
drain valve (46) on the bottom of pressure
regulator (50). The moisture will flow
through vinyl tubing drain (47) on the bottom
of pressure regulator (50) and exit the bot-
tom of MPS enclosure (1).
D