5 measuring in 3-d (cycle 4, fcl 3 function), Cycle run, Please note while programming – HEIDENHAIN 530 (340 49x-07) Cycle programming User Manual
Page 463

HEIDENHAIN iTNC 530
463
1
7.5 MEA
S
URING IN 3-D (Cy
c
le 4, FCL 3 function)
17.5 MEASURING IN 3-D (Cycle 4,
FCL 3 function)
Cycle run
Touch Probe Cycle 4 measures any position on the workpiece in the 
probing direction defined by a vector. Unlike other measuring cycles, 
Cycle 4 enables you to enter the measuring path and feed rate directly. 
Also, the touch probe retracts by a definable value after determining 
the measured value.
1
The touch probe moves from the current position at the entered 
feed rate in the defined probing direction. Define the probing 
direction in the cycle by using a vector (delta values in X, Y and Z).
2
After the TNC has saved the position, the touch probe stops. The 
TNC saves the X, Y, Z coordinates of the probe-tip center (without 
calculation of the calibration data) in three successive Q 
parameters. You define the number of the first parameter in the 
cycle.
3
Finally, the TNC moves the touch probe back by that value against 
the probing direction that you defined in the parameter MB.
Please note while programming:
Cycle 4 is an auxiliary cycle that you can only use in 
conjunction with external software! The TNC does not 
provide any cycle with which you can calibrate the touch 
probe:
The TNC retracts the touch probe by no more than the 
retraction distance MB and does not pass the starting point 
of the measurement. This rules out any collision during 
retraction.
Ensure during pre-positioning that the TNC moves the 
probe-tip center without compensation to the defined 
position!
Remember that the TNC always writes to 4 successive Q 
parameters. If the TNC could not determine a valid touch 
point, the fourth result parameter will have the value –1.
The TNC saves the measured values without calculating 
the calibration data of the touch probe.
With function FN17: SYSWRITE ID 990 NR 6 you can set 
whether the cycle runs through the probe input X12 or 
X13.
