beautypg.com

Misc screen, 3 sof tw a re set u p – HEIDENHAIN ND 1200 User Manual

Page 97

background image

ND 1200 QUADRA-CHEK

97

2.3 Sof

tw

a

re

set

u

p

Misc screen

The Misc screen contains fields for configuring:

„

Optical edge (OE) time-out: The minimum allowed time in
milliseconds between the detection of two edges. This time is
specified to prevent noise from being recognized as valid edges.

„

Optical edge (OE) debounce: The minimum time in milliseconds that
a light transition must be stable to result in a valid edge detection.

To configure edge detection in the Misc screen:

U

Press MENU/SETUP to display the setup menu and then highlight
the Misc menu item.

U

Highlight the OE Time-out (ms) data field and enter the minimum
time between valid edges in milliseconds.

U

Highlight the OE Debounce (ms) data field and enter the minimum
time an edge transition must be stable in milliseconds.

U

Press the FINISH key to save parameters and return to the setup
menu.

Enter OE Time-out

Enter OE Debounce