100

2 Installation and Specifications

2.3 Sof

tw

a

re

set

u

p

Linear error correction (LEC)

LEC compensates for machine irregularities and encoder non-
linearities by applying a single linear correction value to the entire
range of measurement. To apply LEC to a measurement axis:

U

Verify that the crosshair probe is selected. Press the PROBE soft
key to select it if necessary.

U

Press MENU/SETUP to display the setup menu and then highlight
the Scale Factor menu item.

U

Verify that the Active choice field specifies NO.

U

Highlight the SLEC menu item and verify that the Enabled choice
field specifies OFF.

U

Highlight the LEC menu item and make sure that all correction
values are 1.0.

U

Highlight the NLEC setup menu item and verify that the NLEC
choice field specifies OFF.

U

Position the standard artifact along the measurement axis.

U

Align the artifact as closely as possible to the axis and then perform
a skew alignment as described in chapter 1 (see "Align the part to a
measurement axis" on page 34).

When scale factors are used, they must be applied after
error corrections.

LEC can not be performed if a different error correction is
already enabled.

Verify that the Scale Factor is not active

Verify that SLEC compensation is OFF

Verify that NLEC compensation is OFF