Reporting

Reports of measurement results can be sent to a USB printer or USB
flash drive over the USB port, or to a PC over the RS-232 serial port.

Report types include:

Sending reports

Reports can be sent at any time. To send a report:

U

Press the feature SEND key.

Reports can also be sent:

„

By including the send function in a program

„

When the Print setup screen is configured to send when an edge is
recognized

„

When the Measure setup screen is configured to send in response
to tolerance test results

„

By pressing a hot key configured in the Hot Keys setup screen to
send

Report content, destination and format are specified in the
Print, RS-232 and USB setup screens discussed in Chapter
2: Installation and Specifications.

Report type

Report content

Display

The data displayed on the DRO screen is sent as
one row for each axis.

Report

All feature measurement data is sent in tabular
form without tolerance data.

Tol Rpt

All tolerance data is sent in tabular form. Feature
measurement data are not sent.

CSV

All feature measurement data is sent in as comma
separated variables without tolerance data.

Tab

All feature measurement data is sent in as tab
separated variables without tolerance data.

None