beautypg.com

Error messages – HEIDENHAIN ND 760 E User Manual

Page 23

background image

23

To erase error messages:

After you have removed the cause of error:

➤ Press the CL key.

Error Messages

Message

Cause and Effect

SIGNAL X

Encoder signal is too weak,

e.g. when an encoder is

contaminated.

ERR. REF. X

The spacing of the reference

marks as defined in P43 is not

the same as the actual spacing.

FRQ. X

The input frequency for this

encoder input is too high. This

can occur when the scale is

moved too fast.

ERR. MEMORY

Checksum error: Check the

datum, operating parameters and

compensation values for

multipoint axis error

compensation. If the error recurs,

contact your service agency!

Error

Messages

Bateil1.pm6

21.01.2002, 09:39

23