Seaward 350 PrimeTest User Manual
Page 99

PrimeTest 350 Operating
Instructions
- 98 -
15 Electrical
Specifications
15.1 Earth
Continuity
Test Current (2
load)
±200 mA
Test Circuit Voltage
± 4V min
Test Duration
2, 5, 10, 20, 30 seconds
Pass/Fail
Limit
User
Defined
Display
Range
0.00
–
19.99
Measuring Range
0.01 – 19.99
Resolution
0.01
Accuracy
±
5%
±2
digits
15.2 Insulation
Resistance
Test Voltage Open Circuit
250V or 500V +20% / -0%
Test Voltage @ 1mA
250V or 500V min
Test Current Short Circuit <2
mA
Test
Duration 2,
5,
10,
20,
30
seconds
Pass/Fail
Limit
User
Defined
Display
Ranges
0.00
–
1.99M
2.0
–
19.9M
20 – 199M
Measuring
Range
0.01
–
40M
Resolution
0.01M
, 0.00 – 1.99M
0.1M
, 2.0 – 19.9M
1M
, 20 – 199M
Accuracy
± 5% ±2 digits
15.3 Substitute
Leakage
Test Voltage
40V a.c 50Hz open circuit
Test Duration
2s – 30s
Pass/Fail
Limit
User
Defined
Display
Range
0.15
–
19.99mA
Measuring Range
0.25 – 19.99mA
Resolution
0.01mA
Accuracy
± 5% ± 2 digits
Note: The displayed measurements can be scaled
for a supply voltage of 230 V, 244 V or 253 V
depending on configuration( i.e. 230 + 0 %,
+6% or +10%).