Seaward 350 PrimeTest User Manual
Page 100
PrimeTest 350 Operating
Instructions
- 99 -
15.4 Protective Earth Conductor Current
Measurement Method
Differential
Test Voltage
Mains supply 230V +10% / -15%
Max Load Current
16A
Test Duration
2, 5, 10, 20, 30 seconds
Pass/Fail
Limit
User
Defined
Display
Range
0.00
–
19.99mA
Measuring Range
0.25 – 19.99mA
Resolution
0.01mA
Accuracy
±
10%
±2
digits
Attention: Avoid prolonged, repeated use at
full load (16A) and excessive test duration as
this may reduce the life of the unit.
15.5 Touch
Current
Test Voltage
Mains supply 230V +10% / -15%
Max Load Current
16A
Input Impedance
2K
As per IEC 61010-1 A.1
Test Duration
2, 5, 10, 20, 30 seconds
Pass/Fail
Limit
User
Defined
Display
Range
0.00
–
3.50mA
Measuring Range
0.10 – 3.50mA
Resolution
0.01mA
Accuracy
±
10%
±2
digits
Attention: Avoid prolonged, repeated use at
full load (16A) and excessive test duration as
this may reduce the life of the unit.
15.6 Voltage/Frequency
Measurement
Incoming mains voltage and frequency are displayed
during the 2 powered tests only.
Voltage Display Range
0 – 300V
Voltage Measuring Range
20 – 300V
Voltage
Resolution
1V
Voltage
Accuracy
±
10%
Frequency Measuring Range
50Hz or 60Hz
Frequency
Accuracy
Indication
only