PrimeTest 350 Operating

Instructions

- 88 -


The sequence will now progress to the Protective
Conductor Current measurement.

The display now shows the supply voltage and
frequency measured prior to starting the test. The
measured protective conductor current is shown in real
time on both the digital display and the analogue
bargraph.

Note: The vertical bar midway along the analogue

bargraph indicates the Limit value (in this
case 0.75mA).


The test sequence is now complete. The display at the
end of a sequence depends upon the options set in
Menu – User Options – Test Options as described in
section 10.1.3.

The example below shows the end of test screen when
Menu – User Options – Test Options has been
configured for Options Menu.