9200 Plus SCPI Command Summary
Command
Keyword 1
Keyword 2
Keyword 3
Parameter
Range
Notes
:PULSe[0]
Command to change the units’
global settings, this is the same
as using the :SPULse
command
:COUNter
Subsystem. Contains commands
to define the Counter function.
:STATe
0/1 or OFF/ON
Enables/Disables the counter
function.
:SOURce
T
0
, CHA, CHB,
CHC, CHD
Selects the counter source.
Default is T
0
.
:CLear
Clears the active counter.
:COUNt?
Reports the clock pulses output
except, in single shot mode, in
this case the trigger pulses input
on the channel will be displayed.
:CCOunter
0 to 100,000
Cycle counter. Sets the number of
system cycles to repeat when in
duty cycle mode. Setting to zero
disables any cycle counts.
:EXTPeriod
External trigger period.
:STATe
0/1 or OFF/ON
Enables/Disables the counter
function.
:AUTo
0/1 or OFF/ON
Sets auto update mode. Auto will
continually update the value,
over-writing the previous value.
:CLear
Clears the active counter.
:NCOUnts
1 to 15
Number of periods to average.
:COUNt?
Reports the number of clock
counts between input pulses.
:PERiod?
Returns the period value taking
into consideration the number of
averaging counts and system
resolution. Value is in seconds.
:ACTive?
Reports if actively counting
period. State will be zero if not
actively counting the period.
:EXTernal
Submenu for selecting the
trigger input
:MODe
REARM
REARM is Plus additional
feature to the trigger modes.
:UPDate
:AUTO
0/1 or OFF/ON
Turns on or off auto register
update mode. When off registers
are synchronously updated on
command
:EXecute
Initiates a synchronous update if
in manual update (synchronous)