Quantum Composers 9200 Series User Manual
Page 37

9200 Manual Version 1.6 | Quantum Composers
Page 37
:PCOunter
1-1,000,000
Changes the number of on
pulses to output when the
system is in Duty Cycle
mode.
*Note: The commas should
be omitted.
:OCOunter
1-1,000,000
Changes the number of off
pulses to suppress when the
system is in Duty Cycle
mode.
*Note: The commas should
be omitted.
:EXTernal
Submenu for selecting the
trigger input
:MODe
DISabled /
TRIGger /
GATe
Selects the trigger mode*
:LEVel
0.2 - 15[V]
Choose the gate level
threshold to trigger on which
should be set to ~ 50% of the
input voltage.
:EDGe
RISing / FALLing
Selects which edge (rising or
falling) to use as the trigger
signal
:POLarity
LOW / HIGH
Sets the polarity of the gate
signal. HIGH output is active
when the gate signal is high;
LOW output is active when
the gate signal is low.
* It should be noted that when the 9200 is in external trigger mode, a single pulse ONLY will be produced on the
active channels upon each external input trigger. The standard 9200 model cannot be configured to produce a
continuous, burst, or duty cycle pulse train on an external input trigger. The 9200 Plus model may be configured
to output a single pulse, burst of pulses, or duty cycle response of pulses upon each external input. See Appendix
C for more information on the 9200 Plus model.
Command
Keyword 1
Keyword 2
Keyword 3
Parameter Range
Notes
:PULSe[1/2/n]
Command to change the
units’ channel specific
settings.
:STATe
0/1 or OFF/ON
Enables/Disables output
pulse for selected channel.
:WIDTh
10[ns]
– 1000[s]
Sets the pulse width for the
selected channel. The
command should be sent
without units. If for example
50ns is desired the parameter
sent should be 50e-9, or the
decimal equivalent.