Page 50 9200 Manual Version 1.6 | Quantum Composers

Plus Options

Shown below in Figure C.2 is the Plus Options section of the 9200 Plus application followed by
detailed descriptions of the individual functions:

Figure C.2

–9200 Application: System Parameters

Shot Counter:

The “Shot Counter” can be set up to count the system (T

o

) or any of the

available channel outputs. This counter is a 32 bit counter that can be enabled and reset
by the user

by means of the “Enable” selection and “Clear” button.

Period Counter:

The “Period Counter” measures the period of the incoming external

trigger pulses. The period counter must be first

enabled by means of the “Enable”

selection. This counter can average up to 15 periods for greater accuracy. This is
controlled under the “Periods to Average.” If the period counter is selected to run in
manual mode

(“Auto Update” not enabled), it will latch the incoming period measurement

and hold the reading

. If “Auto update” mode is selected, the reading will constantly be

updated as long as external triggers are applied. The displayed measurement may be
cleared at any time

by means of the “Clear” button.

Sync Update:

The ”Sync Update” allows for the channel timers (delays and widths) to be

manually updated so that all changed values are applied synchronously. Normally when
any width or delay is changed, it is automatically applied. When Synchronous Update
Mode is selected to be enabled, all channel width and delays can be adjusted and then
updated simultane

ously via the “Update” button (or SCPI command).

Duty Cycle Counts:

The “Duty Cycle Counts” value sets the number of cycles the duty

cycle sequence will execute. If set at 0, the cycles will continue until stopped by the user.
Again, this feature can be

thought of a “burst” for the duty cycle mode.