9200 SCPI Command Summary

Command

Keyword 1

Keyword 2

Keyword 3

Parameter

Range

Notes

:INSTrument

The units’ upper level command
keyword.

:CATalog

?

Returns a comma separated list
of the names of all channels.
Example: A two channel unit
would return T

0

, CHA, CHB

:FULL

?

Returns a comma-separated list
of the names of all the channels
and their associated number.
Example: a two channel unit
would return T

0

, 0, CHA, 1, CHB,

2

:COMMands

?

Returns an indentured list of all
valid SCPI commands

:NSELect

0

– 4

Selects a channel using the
numeric value

:SELect

T

0

/ CH[A-D]

Selects a channel using the
identifier

:STATe

0/1 or OFF/ON

Enables/Disables the selected
channel output. If no channel
has been selected the command
is applied to T

0.

If T

0

is selected

all outputs are affected.
Enabling T

0

is the same as

pressing the RUN button.


Command

Keyword 1

Keyword 2

Keyword 3

Parameter Range

Notes

:PULSe[0]

Command to change the
units’ global settings, this is
the same as using the
:SPULse command

:STATe

0/1 or OFF/ON

Enables/Disables the output
for all channels. This
command is the same as
pressing the Run/Stop button.

:PERiod

200[ns]

– 1000[s]

Sets the T

0

period. The

command should be sent
without units. If for example
200ns is desired the
parameter sent should be
200e-9, or the decimal
equivalent.

:MODe

NORMal/SINGle/

BURSt/DCYCle

Changes the system output
mode.

:BCOunter

1-1,000,000

Changes the number of
pulses to output when the
system is in burst mode.
*Note: The commas should
be omitted.