beautypg.com

Metro Mobile Computing Workstations MPS-4007 Lithium Ion Nano-Phosphate Power Supply User Manual

Page 13

background image

Metro

TM

MPS-4007 Li-Nano Power Supply Operations Manual Supplement, Doc. L01-505

13

InterMetro Industries Corporation

Equipment Classification

The Metro

TM

MPS-4007 Li-Nano Power Supply is TÜV certified with respect to electric shock,

fire and mechanical hazards only in accordance with UL 60601-1, CAN/CSA 22.2 No. 601.1 and
IEC/EN 60601-1-1

as Type B Patient Care Medical Equipment suitable for use in patient

vicinities.

The MPS-4007 Li-Nano Power Supply is a Class I Equipment that is also internally powered.

According to the degree of protection against electric shock, it is a Type B applied part.

According to the degree of protection against ingress of water the MPS-4007 Li-Nano Power
Supply is considered ordinary.

Warning:

Risk of Explosion

The MPS-4007 Li-Nano Power Supply is not suitable for use in the presence of a
flammable anesthetic mixture with air or with oxygen or nitrous oxide.

The MPS-4007 Li-Nano Power Supply is suitable for continuous operation.

Important Product Notices

The MPS-4007 Li-Nano Power Supply is a subsystem component of the Metro

TM

1700 and

1800 Series Mobile Computing Workstations. Please refer to the respective User Guide for the
Mobile Computing Workstation to which the Power Supply is attached for additional information

.

The MPS-4007 Li-Nano Power Supply is only TÜV certified if the equipment is left in its original
configuration as shipped from InterMetro Industries Corporation. If the user alters the power
supply by changing components on or adding components to the power supply, this will void the
TÜV certification of the power supply and the CSA certification of the Mobile Computing
Workstation and is expressly prohibited by InterMetro Industries Corporation.

The Power Supply used is provided with a “Hospital Grade” or “Hospital Only” attachment plug
for connection to the AC supply circuit.

To ensure equipment grounding reliability, the Power Supply should only be connected to AC
outlet receptacles which are marked “Hospital Grade” or “Hospital Only”. Where the integrity of
the external protective earth connector arrangement is in doubt, equipment shall be operated
from its internal electrical power source.

Disposal

Steel sections of the power supply and LiFePO

4

batteries can be disposed of or recycled.

Electronic components or printed circuit boards, etc., shall be disposed of in accordance with
local, state and federal regulations.

For further information call InterMetro Customer Service.