beautypg.com

Measurement Computing PCI-DAS6071 User Manual

Page 36

background image

PCI-DAS6071 User's Guide

Specifications

36

Table 4. Relative accuracy specifications - all values are (±)

Range

Relative Accuracy (mV)

Single Point

Averaged (Note 2)

±10V

7.37

1.11

±5V

3.68

0.557

±2.5V

1.84

0.278

±1V

0.737

0.111

±500mV

0.368

0.056

±250mV

0.238

0.032

±100mV

0.111

0.015

±50mV

0.082

0.009

0 to 10V

3.68

0.557

0 to 5V

1.84

0.278

0 to 2V

0.737

0.111

0 to 1V

0.368

0.056

0 to 500mV

0.238

0.032

0 to 200mV

0.111

0.015

0 to 100mV

0.082

0.009

Note 2:

Averaged measurements assume dithering and averaging of 100 single-channel readings.

Relative accuracy is defined as the measured deviation from a straight line drawn between measured endpoints
of the transfer function. ADC resolution, noise and front-end non-linearity are included in this measurement.

Table 5. Differential non-linearity specifications

All ranges

±0.5 LSB typ

±1.0 LSB max