Accuracy – Measurement Computing PCI-DAS6071 User Manual
Page 35

PCI-DAS6071 User's Guide
Specifications
35
Accuracy
1.25 MS/s rate, single channel operation and a 15-minute warm-up. Accuracies listed are for measurements
made following an internal calibration. They are valid for operational temperatures within ±1° C of internal
calibration temperature and ±10 °C of factory calibration temperature. Calibrator test source high side tied to
channel 0 high and low side tied to channel 0 low. Low-level ground is tied to channel 0 low at the user
connector.
Table 2. Absolute accuracy specifications
Range
Absolute Accuracy (LSB)
±10V
±2.9
±5V
±2.1
±2.5V
±3.0
±1V
±3.0
±500mV
±3.0
±250mV
±3.1
±100mV
±3.3
±50mV
±3.7
0 to 10V
±2.8
0 to 5V
±4.4
0 to 2V
±4.4
0 to 1V
±4.5
0 to 500mV
±4.6
0 to 200mV
±4.8
0 to 100mV
±5.2
Table 3. Absolute accuracy components specifications
– all values are (±)
Range
% of
Reading
Offset
(mV)
Noise + Quantization (mV)
Temp Drift
(%/DegC)
Absolute
Accuracy at
FS (mV)
Single Pt
Averaged
(Note 1)
±10V
0.0714
6.38
6.10
0.846
0.0010
14.369
±5V
0.0314
3.20
3.05
0.423
0.0005
5.193
±2.5V
0.0714
1.61
1.53
0.211
0.0010
3.605
±1V
0.0714
0.653
0.610
0.085
0.0010
1.452
±500mV
0.0714
0.335
0.305
0.042
0.0010
0.735
±250mV
0.0714
0.176
0.208
0.024
0.0010
0.379
±100mV
0.0714
0.081
0.098
0.011
0.0010
0.163
±50mV
0.0714
0.049
0.071
0.007
0.0010
0.091
0 to 10V
0.0314
3.20
3.05
0.423
0.0005
6.765
0 to 5V
0.0714
1.61
1.53
0.211
0.0010
5.391
0 to 2V
0.0714
0.653
0.610
0.085
0.0010
2.167
0 to 1V
0.0714
0.335
0.305
0.042
0.0010
1.092
0 to 500mV
0.0714
0.176
0.208
0.024
0.0010
0.558
0 to 200mV
0.0714
0.081
0.098
0.011
0.0010
0.235
0 to 100mV
0.0714
0.049
0.071
0.007
0.0010
0.127
Note 1:
Averaged measurements assume dithering and averaging of 100 single-channel readings.
Each PCI-DAS6071 is tested at the factory to assure the board’s overall error does not exceed absolute accuracy
limits described in Table 2.