K-Patents PR-01-S User Manual
Page 94

INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)
DOCUMENT/REVISION No. INM 1/14
Effective: May 15, 2009
92
Figure 11.12
Isolation valve and retraction unit HIMP-2, standard delivery.
The Retractor HIMP-2 is delivered according to Figure 11.12 with an isolation valve (a ball valve with 
handle), a stuffing box and threaded force bars with covers. The following components are also included in 
a standard delivery: 
1. steam wash nozzle 
2. three ball valves (one for wash nozzle, one for blow out valve, one for flushing the box) 
3. two check-valves (one for wash nozzle and one for stuffing box) 
4. piping between steam wash line and stuffing box 
5. steam input connection R 1/4” 
The Retraction unit HIMP-2 is used in combination with a special version of the refractometer 
PR-01-S-LPH only, Figure 11.13. The process connection is a Sandvik clamp. 
Figure 11.13
Model PR-01-S-LPH, Long Probe adapted to Retraction unit HIMP-2.