K-Patents PR-01-S User Manual
Page 31
INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)
DOCUMENT/REVISION No. INM 1/14
Effective: May 15, 2009
29
3
2
2
3
7
5
SD
RD
SGND
SGND DAT- DAT+
Receive data
Transmit data
Signal ground
INDICATING TRANSMITTER
D25 D9
DCE
Figure 3.70
RS-232 connection to DCE-type equipment (e.g., modem).
2
3
4
7
8
1
3
2
5
8
20 4
7
5
6
6
Transmit data
Receive data
Signal ground
INDICATING TRANSMITTER
Request to send
Clear to send
Data set ready
Data carrier detect
Data terminal ready
SD
RD
SGND
SGND DAT- DAT+
D25 D9
DTE
Figure 3.71
RS-232 connection to DTE-type equipment (e.g., computer).
The character structure conforms to the ISO 1177 standard. It is compatible with the 'RS-232' interfaces of
personal computers.
The character parameters are configuration selectable:
speed 1200 to 9600 bits/s
parity odd, even or none
stop bits 1 or 2
flow control hardware, XON / XOFF or none.
The set of measurement results is output in a fixed-length record. The record consists of variable-length
numeric fields at fixed locations in the record. The gaps are filled with ASCII spaces (code 20 hex).
The record format is (program version 4.0 and newer):
Column Field
Format
Value
1
TEST
float, 1 decimal
raw refraction value
16
CONC%
float, selected decimals concentration
value
31
temp C
float, 1 decimal
temperature, degrees centigrade
46
RI 25 C
float, 4 decimals
standard RI at 25 C
61
phase
1 digit
operation phase code
63 error integer
error
code
66
divert
integer
divert control status code
68 checksum
integer
76
terminator
CR and LF