beautypg.com

K-Patents PR-01-S User Manual

Page 49

background image

INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)

DOCUMENT/REVISION No. INM 1/14

Effective: May 15, 2009

47

Code

Diagnostic message

Cause

0

Normal operation

No fault

0

In divert control operation

No fault

0

Removed from divert control

No fault

Process related faults

Priority Code

Diagnostic message

Cause

1

1

Retrying wash

Prism wash attempt failed

2

2

Solids warning

Solids below warning limit

3

3

Solids alarm

Solids below alarm limit

4

23

Wash stop / Temp. limit

Temperature below set limit

5

4

Low concentration / no sample

Whole image bright

6

6

High process temperature

Temperature above set limit

7

21

Low light intensity

Image nearly dark

8

22

Dip in image

Optical image not correct

9

7

High conc / prism coated

No shadow edge in image

Equipment malfunctions

Priority Code

Diagnostic message

Cause

10

28

Wash stop / ext. stop

Input switch activated for wash stop

11

5

Prism wash failure

Wash fault after retries

12

8

External output failure

No response from output unit

13

9

Relay unit fault

No response from relay unit

14

10

Divert control fault

No response from divert unit

15

11

Temperature measurement fault

Temperature out of limits

16

12

Prism coated / LED fault

Light intensity too low

17

13

Not in line

In line switch contacts open

18

24

High internal humidity

Sensor internal humidity too high

19

25

High internal temperature

Sensor internal temp too high

20

14

Detector timeout

No signal from image detector

21

15

Sensor transmission error

Noisy signal from sensor

22

16

No sensor signal

No signal from sensor

23

17

Wrong concentration parameters Makes R.I. outside range

24

18

Constants error

Wrong EPROM parameters

25

19

Sensor interface fault

Fault on processor card

26

20

EEPROM write error

Fault on processor card

Phase codes

Code

Diagnostic message

Cause

0 Blank

Normal

operation

1

Wait

Wash permit wait

2 Precondition Preconditioning
3

Precondition

Precondition / wash wait

4 Wash

Wash

5

Recovery

Recovery from wash

6

Hold

Hold after external wash


Table 6.30