INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)
DOCUMENT/REVISION No. INM 1/14
Effective: May 15, 2009
32
3.8. FUSES
Fuses printed on circuit board PR-7030:
Fuse F1, F2: 5 x 20 mm, T1A (slow)
AC Main power protects electronics against wrong
primary voltage
Fuse F3: 5 x 20 mm, T63 mA (slow)
4-20 mA output protection
Fuse F4: 5 x 20 mm, T2A (slow)
Secondary main fuse
Fuse F5: 5 x 20 mm, T500 mA (slow)
Sensor power protection
Fuse F6: 5 x 20 mm, T500 mA (slow)
Serial bus protection
Fuse F7: 5 x 20 mm, T1A (slow)
Processor card protection
Note. For a CSA-Certified instrument use only CSA-Certified fuses F1, F2, F4-F7.
Fuses for +24 V DC main power
Fuses printed on circuit board PR-7030 when IT-R has +24V DC input voltage
Fuse F1, F2: 5 x 20 mm, T2A (slow)
DC Main power protects electronics against wrong
primary voltage
Fuse F3: 5 x 20 mm, T63 mA (slow)
4-20 mA output protection
Fuse F4: 5 x 20 mm, T2A (slow)
Secondary main fuse
Fuse F5: 5 x 20 mm, T500 mA (slow)
Sensor power protection
Fuse F6: 5 x 20 mm, T500 mA (slow)
Serial bus protection
Fuse F7: 5 x 20 mm, T1A (slow)
Processor card protection