K-Patents PR-01-S User Manual
Page 84
![background image](https://www.manualsdir.com/files/793662/content/doc084.png)
INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)
DOCUMENT/REVISION No. INM 1/14
Effective: May 15, 2009
82
8.5. RECOMMENDED WASH PRESSURES AND TIMES
To select a recommended wash pressure use the following table:
Nozzle type
Integral nozzles
Wash medium Wash time
seconds
Minimum above
process pressure
Maximum above
process pressure
- HPS
steam
2-3
2 bar (30 psi)
6 bar (85 psi)
- HPN
pressure water
10-15
30 bar (450 psi)
70 bar (1000 psi)
HIMP-2
steam
3-5
2 bar (60 psi)
6 bar (85 psi)
HIMP-3
water
10-15
30 bar (450 psi)
70 bar (1000 psi)
Flow cell nozzles
PR-3356 / WP
pressure water
10-15
30 bar (450 psi)
70 bar (1000 psi)
PR-3355 / SN
steam
3
2 bar (30 psi)
4 bar (70 psi)
PR-3354 / WN
water
10
2 bar (30 psi)
4 bar (70 psi)
Table 8.50
Recommended wash pressures and times.
Note. Steam wash:
Do not use longer wash times than is recommended in the Table 8.50. In case of
coating adjust the wash interval.