K-Patents PR-01-S User Manual
Page 33

INSTRUCTION MANUAL FOR K-PATENTS PR-01-S (-AX/FM/CS)
DOCUMENT/REVISION No. INM 1/14
Effective: May 15, 2009
31
Error status: For program versions from 4.0 the error status codes are:
Code
Priority
Error status
0 0 No
errors
Process disturbances:
1 1 Retrying
wash
2 4 Solids
warning
3 5 Solids
alarm
4
7
Low concentration / no sample
5
13
Prism wash failure
Malfunctions:
6
8
High process temperature
7
11
High conc / prism coated
Instrument malfunctions:
8
14
External output fault
9
15
Relay unit fault
10
16
Divert control fault
11
17
Temperature measurement fault
12
18
Prism coated / LED fault
13
19
Not in line
14 20 Detector
timeout
15
21
Sensor transmission error
16
22
No sensor signal
17
23
Wrong concentration parameters
18 24 Constants
error
19
25
Sensor interface fault
20 26 EPROM
fault
From version 5.0, process disturbances
21
9
Low light intensity
22
10
Dip in image
23
6
Wash stop / Temp. limit
From version 6.0, process disturbances:
24
2
High internal humidity
25
3
High internal temperature
From version 6.0, instrument malfunctions:
26
12
Wash stop / ext. stop
Checksum: The checksum is the arithmetic sum of ASCII codes in columns before the checksum field.
For programs until version 3.1 the range is columns 1 to 65. For programs from version 4.0 the range is
columns 1 to 67. Least significant 7 bits are taken into the sum: the parity bits are zeroed.
Record terminator: The record is terminated with ASCII characters CR (0d hex) and LF (0a hex) to make
the record a text line.