Error status: For program versions from 4.0 the error status codes are:

Code

Priority

Error status

0 0 No

errors

Process disturbances:

1 1 Retrying

wash

2 4 Solids

warning

3 5 Solids

alarm

4

7

Low concentration / no sample

5

13

Prism wash failure

Malfunctions:

6

8

High process temperature

7

11

High conc / prism coated

Instrument malfunctions:

8

14

External output fault

9

15

Relay unit fault

10

16

Divert control fault

11

17

Temperature measurement fault

12

18

Prism coated / LED fault

13

19

Not in line

14 20 Detector

timeout

15

21

Sensor transmission error

16

22

No sensor signal

17

23

Wrong concentration parameters

18 24 Constants

error

19

25

Sensor interface fault

20 26 EPROM

fault

From version 5.0, process disturbances

21

9

Low light intensity

22

10

Dip in image

23

6

Wash stop / Temp. limit

From version 6.0, process disturbances:

24

2

High internal humidity

25

3

High internal temperature

From version 6.0, instrument malfunctions:

26

12

Wash stop / ext. stop

Checksum: The checksum is the arithmetic sum of ASCII codes in columns before the checksum field.
For programs until version 3.1 the range is columns 1 to 65. For programs from version 4.0 the range is
columns 1 to 67. Least significant 7 bits are taken into the sum: the parity bits are zeroed.

Record terminator: The record is terminated with ASCII characters CR (0d hex) and LF (0a hex) to make