beautypg.com

6 appendix f - specifications – Guralp Systems CMG-CD24S6EAM User Manual

Page 41

background image

CMG-CD24S6EAM

Appendices

4.6

Appendix F - Specifications

Parameter

Specification

Input range

±10V differential

Nominal Sensitivity

0.9µV/count

Standard output format

24-bit

Noise-free resolution (NPR) at 20

samples/s

> 132 dB r.m.s. (> 22 bits)

Digital signal processor

TMS3200 at 144 MHz

Output latency

<100ms at 100sps (low latency port only)

Output baud rate

115200

Operating temperature range

–10 to +75 °C

Internal thermometer accuracy

±0.33 °C (30 °C)

±0.5 °C (10..50 °C)
±1.0 °C (–10..75 °C)

Internal thermometer linearity

±0.5 °C

Internal thermometer resolution

0.0625 °C

Dimensions
(excluding connectors)

483 x 280 x 89mm (excluding connectors)

Weight

1.95kg

Voltage requirements

10..24V DC

Current at 12 V DC with GPS

270 mA

MAN-C24-0003

38

Issue A - May 2011