beautypg.com

Guralp Systems CMG-CD24S6EAM User Manual

Page 28

background image

CMG-CD24S6EAM

Appendices

2.

If the digitiser is to be connected to a routed network, enter the address
of the default gateway in the same manner.

3.

At the netmask prompt, enter 0 if you are using normal (classful) internet

addressing. If you have a non-standard or classless addressing scheme,
enter the number of bits of the IP address to be used for network

identification. Note that this is not the standard way of specifying
netmasks; it is more akin to the /n modifier used in CIDR address

specification.

4.

When you have finished setting up the module, apply the new settings
by selecting option 9 Save and Exit. The module will re-boot with the

new settings in effect.

5.

Disconnect the power supply from the acquisition unit.

6.

Remove the RS232 cable from the 9-pin D-connector.

7.

Connect the 26-way IDE-connector to the digitiser.

8.

Install the top cover.

9.

Connect the sensors to the acquisition unit.

10. Connect the power supply to the acquisition unit.

MAN-C24-0003

25

Issue A - May 2011