Pacemaker waveform, Pacemaker amplitude, Pacemaker width – Fluke Biomedical MPS450 User Manual
Page 90

MPS450
Operators Manual
4-8
Pacemaker Waveform
Table 4-8 lists the numeric control codes for the pacemaker waveform functions.
Table 4-8. Numeric Codes for Pacemaker Waveform Actions
Numeric
Control Code
Action
Remote Control
Entry
383
Run atrial pacer wave.
ATR
110
Run asynchronous pacer wave.
ASN
111
Run demand frequent sinus pacer wave.
DFS
112
Run demand occasional sinus pacer wave.
DOS
113
Run AV sequential pacer wave.
AVS
114
Run noncapture pacer wave.
NCA
115
Run nonfunction pacer wave.
NFU
Pacemaker Amplitude
Table 4-9 lists the numeric control codes for the pacemaker amplitude functions.
Table 4-9. Numeric Codes for Pacemaker Amplitude Actions
Numeric
Control Code
Action
Remote Control
Entry
384
Set pacer amplitude to 1 mV.
PA1
385
Set pacer amplitude to 2 mV.
PA2
386
Set pacer amplitude to 5 mV.
PA5
242, 387
Set pacer amplitude to 10 mV.
PA10
Pacemaker Width
Table 4-10 lists the numeric control codes for the pacemaker width functions.
Table 4-10. Numeric Codes for Pacemaker Width Actions
Numeric
Control Code
Action
Remote Control
Entry
243
Set pacer width to 0.1 ms.
PW0.1
244
Set pacer width to 0.5 ms.
PW0.5
245
Set pacer width to 1.0 ms.
PW1.0
246
Set pacer width to 1.5 ms.
PW1.5
247
Set pacer width to 2.0 ms.
PW2.0