Adding respiration artifact to the bp signal – Fluke Biomedical MPS450 User Manual

Page 64

background image

MPS450

Operators Manual

2-32

gje031.eps

Alternatively, in the Numeric-Control Mode, press the number keys for a dynamic-wave

setting according to Table 2-25, and select RUN:

Table 2-25. Numeric Codes for Dynamic-Wave Settings

Dynamic-Wave Setting

NumericCode

BP1 Arterial 120/80

060

BP1 Radial Art 120/80

061

BP1 Left Vent 120/00

062

BP1 Right Vent 25/00

063

BP2 Arterial 120/80

070

BP2 Radial Art 120/80

071

BP2 Left Vent 120/00

072

BP2 Right Vent 25/00

073

BP2 Pulmonary Art 25/10

074

BP2 Pulm Art Wedge 10/2

075

BP2 Left Atrium 14/4

076

BP2 Right Atrium CVP 15/10

077

BP3 Arterial 120/80

080

BP3 Radial Art 120/80

081

BP3 Left Vent 120/00

082

BP3 Right Atrium CVP 15/10

083

BP3 Pulmonary Art 25/10

084

BP3 Pulm Art Wedge 10/2

085

BP3 Left Atrium 14/4

086

BP3 Right Vent 25/00

087

Adding Respiration Artifact to the BP Signal

The MPS450 adds respiration artifact to any BP-channel dynamic wave, including the

Swan-Ganz-wave series. The artifact is turned on separately for each of the four BP

channels.
The following peak-to-peak changes (mmHg) are caused by the respiration wave, with

values independent of the programmed respiration amplitude: