Setting r-wave width – Fluke Biomedical MPS450 User Manual
Page 57

Cardiac Functions
ECG Testing
2
2-25
screen is active when displayed and remains active until the setting is changed.
Alternatively, in the Numeric-Control Mode, press the number keys for R-wave rate
according to Table 2-19, and select RUN:
Table 2-19. Numeric Codes for R-Wave-Rate Settings
R-Wave-Rate Setting
Numeric Code
30 BPM
312
60 BPM
313
80 BPM
314
120 BPM
315
200 BPM
316
250 BPM
317
Setting R-Wave Width
The MPS450 offers twenty-two preconfigured settings (ms) for R-wave width: nineteen
settings ranging from 20 to 200 ms (10 ms steps), plus three settings at 8, 10, or 12 ms.
Action in the Menu-Control Mode
1. Press the top-menu key labeled PERF.
2. Select RWDET.
3. Select SEL to toggle to the screen for adjusting R-wave width (WIDTH/ampl), with
WIDTH in upper-case letters.
4. Select DOWN or UP to scroll to the desired R-wave width.
5. The width identified on the LCD screen is active when displayed and remains active
until the setting is changed.
Alternatively, in the Numeric-Control Mode, press the number keys for R-wave width
according to Table 2-20, and select RUN:
Table 2-20. Numeric Code R-Wave Width Settings
R-Wave-Width Setting
Numeric Code
8 ms
318
10 ms
319
12 ms
320
20 ms
321
30 ms
322
40 ms
323
50 ms
324
60 ms
325
70 ms
326
80 ms
327
90 ms
328