Cardiac-Output Wave / Injectate

Table 4-40 lists the numeric control codes for the BP cardiac-output wave / injectate

functions.

Table 4-40. Numeric Codes for Cardiac-Output Wave / Injectate Actions

Numeric

Control Code

Action

Remote Control

Entry

(N/A)

Run CO wave 2.5 l/min.

COW2.5

(N/A)

Run CO wave 5.0 l/min.

COW5.0

(N/A)

Run CO wave 10.0 l/min.

COW10.0

(N/A)

Run CO wave faulty injectate.

COWFLT

(N/A)

Run CO wave left/right shunt.

COWLRS

(N/A)

Run CO wave cal pulse.

COWCAL

(N/A)

Set CO injectate to 0 degrees C.

COI0

(N/A)

Set CO injectate to 24 degrees C.

COI24

090

Turn CO wave off.

CRSET

091

Run CO wave 0 degrees C at 2.5 l/min.

(N/A)

092

Run CO wave 0 degrees C at 5.0 l/min.

(N/A)

093

Run CO wave 0 degrees C at 10.0 l/min.

(N/A)

094

Run CO wave 24 degrees C at 2.5 l/min.

(N/A)

095

Run CO wave 24 degrees C at 5.0 l/min.

(N/A)

096

Run CO wave 24 degrees C at 10.0 l/min.

(N/A)

097

Run CO wave faulty injectate 0 degrees C.

(N/A)

099

Run CO wave faulty injectate 24 degrees C.

(N/A)

098

Run CO wave left/right shunt 0 degrees C.

(N/A)

100

Run CO wave left/right shunt 24 degrees C.

(N/A)

101

Run CO wave cal pulse 0 degrees C.

(N/A)

102

Run CO wave cal pulse 24 degrees C.

(N/A)

FHR Rate (Fixed)

Table 4-41 lists the numeric control codes for the FHR rate (fixed) functions.

Table 4-41. Numeric Codes for FHR Rate (Fixed) Actions

Numeric

Control Code

Action

Remote Control

Entry

409

Run fetal ECG at 60 BPM.

F60

410

Run fetal ECG at 90 BPM.

F90

411

Run fetal ECG at 120 BPM.

F120

412

Run fetal ECG at 140 BPM.

F140