beautypg.com

Ventricular tachycardia, Ventricular fibrillation – Fluke Biomedical MPS450 User Manual

Page 51

background image

Cardiac Functions

Arrhythmia Functions

2

2-19

Table 2-13. Numeric Codes for Multiple-PVC Settings

Multiple-PVC Setting

Numeric Code

PAIR PVCs

027

RUN 5 PVCs

028

RUN 11 PVCs

029

Ventricular Tachycardia

A faster-than-normal rhythm of beats (160 BPM) originating in the ventricles, similar to

type-1 (left-focus) PVCs.
Ventricular tachycardia is a life-threatening arrhythmia in which one or multiple, ectopic,

ventricular pacemakers in the bundle branches, Purkinje network, or ventricular

myocardium are firing in a heart beating more frequently than 110 times a minute. In

some cases the heart will be beating at a rate above 240 BPM. Ventricular tachycardia

usually occurs in cases of extreme cardiac disease and often initiates or degenerates into

ventricular fibrillation. This type of tachycardia can reduce cardiac output by as much as

25 % due, in many cases, to the lack of an atrial “kick” and therefore the lack of a

complete filling of the ventricles with blood prior to ventricle contraction.
The MPS450 simulates a ventricular tachycardia at 160 BPM, with beats similar to type-1

(left-focus) PVCs.

Action in the Menu-Control Mode

1. Press the top-menu key labeled ARRHY.
2. Select VENT.
3. Scroll to the LCD screen VENTRICULAR TACH.
4. Select RUN. The ventricular-tachycardia simulation runs continuously (repeats) until

another arrhythmia selection is made.

Alternatively, in the Numeric-Control Mode, press the number keys 030, and select

RUN.

Ventricular Fibrillation

An irregular ventricular waveform, coarse or fine.
Coarse and fine ventricular fibrillations occur when the electrical signals in the ventricles

are chaotic, and multiple, ectopic, ventricular pacemakers are firing erratically. There are

no real P waves and no clear R-R interval. Ventricular fibrillation waveforms are

irregularly shaped. Ventricular fibrillation is a life-threatening condition; usually in such

situations a defibrillator is applied immediately to return the heart to its normal rhythm.
The MPS450 simulates irregular ventricular waveforms; the amplitude of the ventricular

signal is higher for coarse, and lower for fine, fibrillation.

Action in the Menu-Control Mode

1. Press the top-menu key labeled ARRHY.
2. Select VENT.
3. Scroll to the LCD screen for the desired simulation: VENT FIB COARSE or VENT

FIB FINE.

4. Select RUN. The selected ventricular-fibrillation simulation runs continuously

(repeats) until another arrhythmia selection is made.