Cardiac Functions
Fetal / Maternal ECG (Option)
2
2-41
Action in the Menu-Control Mode
1. Press the top-menu key labeled MORE.
2. Select FETAL MATR.
3. Select PER to scroll to a different contraction-period setting.
4. Select IUP to scroll to the desired wave simulation: EARLY DEC (early
deceleration); LATE DEC (late deceleration); or ACC (acceleration).
5. Select RUN. On the LCD screen, RUNNING flashes while the simulation is active.
6. A manually generated IUP ends after one contraction. To repeat the simulation, select
RUN again.
7. Periodically generated IUPs repeat at the selected period. To terminate a periodic-
IUP simulation, select END.
Alternatively, in the Numeric-Control Mode, press the number keys for an IUP-wave
simulation using one of the codes in Table 2-33, and select RUN. (The selected IUP is
generated according to the specific period that was last selected, whether manual, or once
every 2, 3, or 5 minutes.)
Table 2-33. Numeric Codes for IUP Wave Settings
IUP Wave Setting
Numeric Code
IUP OFF
400
IUP EDEC (early deceleration)
402
IUP LDEC (late deceleration)
403
IUP UACC (acceleration)
404
To select a contraction period in the Numeric-Control Mode, press the number keys for
one of the settings found in Table 2-34, and select RUN. (The selected period is applied
to a run of the specific IUP-wave simulation that was last selected, whether early
deceleration, late deceleration, or acceleration.)
Table 2-34. Numeric Codes for IUP Contraction Period Settings
IUP Contraction Period Setting
Numeric Code
IUP MANUAL
405
IUP 2 MIN
406
IUP 3 MIN
407
IUP 5 MIN
408