Connection the mps450 – Fluke Biomedical MPS450 User Manual

Page 22

background image

MPS450

Operators Manual

1-8

Connection the MPS450

gje041.eps

Figure 1-1. MPS450 Patient Simulator

The right side of the MPS450 features connections for linking to blood-pressure, cardiac-

output, and temperature monitors. In addition, there is an auxiliary connection for future

expansion. The bottom of the MPS450 features an RS-232 serial port and a connection

for a battery eliminator. Prewired cables compatible with all major manufacturers’

monitors are available for simulating functions related to blood pressure, temperature,

and cardiac-output injectate. (Call your sales representative for a complete list.)
The left side of the MPS450 features a full set of universal ECG jacks, enabling the

connection of any 3-, 5-, or 12-lead ECG device. AHA and IEC color-coded dots run

along the left side of the face of the unit as an aid in connecting the corresponding U.S.

and international patient leads to the proper universal ECG jacks on the MPS450:

Table 1-2. ECG Jack Labeling

Label Meaning

RA or R

Right arm

LA or L

Left arm

RL or N

Right leg (reference or ground)

LL or F

Left led

V1, V2, V3, V4, V5, and V6

V Leads (U.S. and Canada). Also referred to as
pericardial, precordial, or unipolar chest leads.

C1, C2, C3, C4, C5, and C6

Chest leads (International)