Ecg artifact stimulation – Fluke Biomedical MPS450 User Manual

Page 89

background image

Remote Operations

Codes and Actions

4

4-7

Table 4-6. Numeric Codes for STE Elevation Action (cont.)

Numeric

Control Code

Action

Remote Control

Entry

236

Set ST elevation to -0.7 mV.

STD-0.7

235

Set ST elevation to -0.6 mV.

STD-0.6

234

Set ST elevation to -0.5 mV.

STD-0.5

233

Set ST elevation to -0.4 mV.

STD-0.4

232

Set ST elevation to -0.3 mV.

STD-0.3

231

Set ST elevation to -0.2 mV.

STD-0.2

230

Set ST elevation to -0.1 mV.

STD-0.1

221

Set ST elevation to -0.05 mV.

STD-0.05

220

Set ST elevation to 0.

STD0

219

Set ST elevation to +0.05 mV.

STD+0.05

229

Set ST elevation to +0.1 mV.

STD+0.1

228

Set ST elevation to +0.2 mV.

STD+0.2

227

Set ST elevation to +0.3 mV.

STD+0.3

226

Set ST elevation to +0.4 mV.

STD+0.4

225

Set ST elevation to +0.5 mV.

STD+0.5

224

Set ST elevation to +0.6 mV.

STD+0.6

223

Set ST elevation to +0.7 mV.

STD+0.7

222

Set ST elevation to +0.8 mV.

STD+0.8

ECG Artifact Stimulation

Table 4-7 lists the numeric control codes for the ECG artifact Stimulation functions.

Table 4-7. Numeric Codes for ECG Artifact Stimulation Actions

Numeric

Control Code

Action

Remote Control

Entry

104

Set ECG artifact to off.

EAOFF

105

Set ECG artifact to 50 Hz sine.

EA50

106

Set ECG artifact to 60 Hz sine.

EA60

107

Set ECG artifact to muscle.

EAMSC

108

Set ECG artifact to wandering baseline.

EAWNDR

109

Set ECG artifact to respiration.

EARESP