Adjusting the respiration rate – Fluke Biomedical MPS450 User Manual

Page 78

background image

MPS450

Operators Manual

3-4

Alternatively, in the Numeric-Control Mode, press the number keys for a respiration-lead

setting according to Table 3-1, and select RUN:

Table 3-1. Numeric Codes for Respiration-Lead Settings

Respiration-Lead Setting

Numeric Code

RESP LEAD LA

389

RESP LEAD LA STORED

390

RESP LEAD LL

391

RESP LEAD LL STORED

392


For baseline impedance, press the number keys for a respiration-baseline setting

according to Table 3-2, and select RUN:

Table 3-2. Numeric Codes for Respiration-Baseline Settings

Respiration-Baseline Setting

Numeric Code

RESP BASE 500 OHM

180

RESP BASE 500 STRD

393

RESP BASE 1000 OHM

181

RESP BASE 1000 STRD

394

RESP BASE 1500 OHM

182

RESP BASE 1500 STRD

395

RESP BASE 2000 OHM

183

RESP BASE 2000 STRD

396

Adjusting the Respiration Rate

Respiration rate is measured by the number of times per minute the lungs inflate

(inspiration) and deflate (expiration), expressed as breaths-per-minute (BrPM). Pediatric

respiration rates are two or three times higher than adult rates. Newborn babies, for

example, breathe from forty to fifty times per minute, while adults typically breathe

approximately fifteen times per minute.
The MPS450 simulates nine preconfigured respiration rates (BrPM): 15, 20, 30, 40, 60,

80, 100, and 120, plus a respiration-off condition.

Action in the Menu-Control Mode

1. Press the top-menu key labeled RESP. If the respiration-rate setting has not been

changed, the screen displays the default setting (20 BRPM):

gje024.eps