beautypg.com

Fluke Biomedical MPS450 User Manual

Page 67

background image

Cardiac Functions

Blood Pressure Function

2

2-35

4. As the automatic simulation proceeds, the LCD screen continues to indicate which

wave in the cycle is currently running: RIGHT ATR CVP 15/10; RIGHT VENT

25/00; PULM ARTERY 25/10; or PULM WEDGE 10/2.

5. To terminate the automatic Swan-Ganz simulation, press the ESC key.
6. To generate the complete Swan-Ganz cycle manually, select MAN from the screen

for the BP4 channel to display the screen for BP4 SWAN-GANZ MANUAL:

gje036.eps

7. The manual simulation begins with a right-atrium CVP 15/10 dynamic wave. Select

INSRT to start the Swan-Ganz procedure with the simulated insertion of the catheter

into the heart.

8. To complete the Swan-Ganz procedure manually, select INSRT again. Next, select

INFL (simulating balloon inflation), then DEFL (simulating balloon inflation).

Finally, select PLBK (simulating pullback, that is, the withdrawal of the catheter),

then PLBK again.

9. At each point in the procedure, the LCD screen identifies which dynamic wave in the

cycle of four is currently running.

10. Press the ESC key to return to the BP4 screen.
11. Press the ESC key again to return to the top menu.
Alternatively, in the Numeric-Control Mode, press the number keys for a Swan-Ganz

simulation according to Table 2-27, and select RUN:

Table 2-27. Numeric Codes for Swan-Ganz-Simulation Settings

Swan-Ganz-Simulation Setting

Numeric Code

Swan-Ganz Auto

088

BP4 Right Atrium CVP 15/10

378

BP4 Right Vent 25/00

379

BP4 Pulmonary Art 25/10

380

BP4 Pulm Art Wedge 10/2

381

Swan-Ganz Manual

416

Insert (in Swan-Ganz Manual)

417

Inflate (in Swan-Ganz Manual)

418

Deflate (in Swan-Ganz Manual)

419

Pull Back (in Swan-Ganz Manual)

420