Temperature, Cardiac output, Fetal / maternal-ecg – Fluke Biomedical MPS450 User Manual
Page 31: Computer setup

Introduction
and
Specifications
Detailed Specifications
1
1-17
Right atrium (central venous or CVP): 15/10
Right ventricle: 25/00
Pulmonary artery: 25/10
Pulmonary-artery wedge: 10/2
Left atrium: 14/4
Dynamic Waveforms, p4 ...................................... Swan-Ganz sequence:
Right atrium (CVP)
Right ventricle RV)
Pulmonary artery (PA)
Pulmonary-artery wedge (PAW)
Respiration Artifact............................................... BP delta changes from 3 to 16 mmHg
BP Output .............................................................. Mini DIN 7-Pin
Power-On Default.................................................. 0 mmHg, transducer sensitivity 5
μ
V/V/mmHg
Temperature
Temperature .......................................................... 0
°
C (42
°
F), 24
°
C (75.2
°
F), 37
°
C (98.6
°
F), and 40
°
C (104
°
F)
Accuracy................................................................
±
0.1
°
C
Compatibility ......................................................... Yellow Springs, Inc. (YSI) Series 400 and 700
Output .................................................................... mini DIN 7-pin
Power-On Default.................................................. 0
°
C (42
°
F)
Cardiac Output
Catheter Type ........................................................ Baxter Edwards, 93a-131-7f
Calibration Coefficient.......................................... 0.542 (0
°
C injectate), 0.595 (24
°
C injectate)
Blood Temperature ............................................... 37
°
C (98.6
°
F)
±
2 %
Injectate Volume ................................................... 10 cc
Injectate Temperature........................................... 0
°
C or 24
°
C
±
2 % value
Cardiac Output ...................................................... 2.5, 5, 10 liters per minute
±
5 %
Faulty-Injectate Curve .......................................... Waveform for simulation available
Left-to-Right-Shunt Curve.................................... Waveform for simulation available
Calibrated Pulse.................................................... 1.5
°
for 1 second (37
°
→
35.5
°
)
Repeatability..........................................................
±
1 %
Power-On Default ................................................. 2.5 liters per minute, 0
°
C injectate
Fetal / Maternal-ECG
Fetal Heart Rate (Fixed)........................................ 60, 90, 120, 140, 150, 210, and 240 BPM
Fetal Heart Rate (IUP): .......................................... 140 BPM at beginning, then varies with pressure
Intrauterine-Pressure Waveforms ....................... Early deceleration, late deceleration, and acceleration
Wave Duration....................................................... 90 seconds, bell-shaped pressure curve, from 0 to 90 mmHg and
returning to 0
IUP Period.............................................................. 2, 3, or 5 minutes; and manual
Power-On Default.................................................. FHR 120 BPM, late deceleration, manual
Computer Setup
Port......................................................................... Bidirectional (Data Communications Equipment) RS-232
Baud Rate .............................................................. 9600
Parity ...................................................................... None
Stop Bits ................................................................ 1
Data Bits ................................................................ 8