beautypg.com

4 install system, 5 leak check, 6 electrical connection – Bronkhorst FLOW-SMS Series User Manual

Page 7

background image

BRONKHORST HIGH-TECH B.V.

2.4 Install system

9.17.057

page 7

Install the FLOW‐SMS assembly in accordance with the 
directions following the schematic diagram
Tighten the fittings according to the instructions.  
For compression type adapters, slide the nut and the ferrule 
set in the right order at the tube. Insert the tubing fully into 
the Adapter plate against the shoulder. Rotate the nut finger‐ 
tight, while holding the fitting body steady. Tighten the nut 
three quarters clockwise for the fittings up to 4 mm and 
3/16” tube OD. For larger fittings, tighten the nut one and one‐quarter. 
For metal and o‐ring face seal type, rotate the nut finger‐tight, than turn the nut one‐quarter 

clockwise, which is sufficient for the sealing.  When the Adapter Plate is replaced the complete 
module has to be tested for leakage and function.
 

The preferred mounting position of mass flow or pressure meters/controllers is horizontal. Other 
mounting positions may introduce a zero shift and/or little gas and pressure dependency of the zero 
signal. When mounting the assembly other than horizontal, zeroing of the instruments is advised. The 
zeroing procedure is described in paragraph 2.10. 

Avoid installation in close proximity of mechanic vibration and/or heat sources. 

2.5 Leak check

Check the system for leaks before applying (fluid) pressure. Especially if toxic, explosive or other 
dangerous fluids are used! 

2.6 Electrical connection

Electrical connections must be made with standard cables or according to the applicable hook‐up 
diagrams
. These documents can be found at: 

http://www.bronkhorst.com/en/downloads

 

Please note that FLOW‐SMS instruments are rated IP40, implying that the electronics housings and 
electrical connections do not offer any protection against moist environments. 

The instruments contain electronic components that are susceptible to damage by electro static 
discharge
. Proper handling procedures must be taken during installation, removing and connecting the 
electronics. 

All instruments described in this manual carry 
the CE‐mark and are complient with the EMC 
requirements.
 However compliance with the 
EMC requirements is not possible without the 
use of proper cables and connector/gland  
assemblies. For good results Bronkhorst High‐ 
Tech B.V. can provide standard cables. 
Otherwise follow the guidelines as stated 
hereby.