2.13.11

Blind/Flow plates

-

Removal 
Ensure that the system is depressurised before dismounting  
the blind/flow plates. Remove the Blind/Flow plates assy by 
loosening the 4 bolts counter clockwise to release the assy.  

9.17.057

page 16

FP‐XX  

 

BP‐XX   

 

 
-

Replacement 
Replace the gaskets AS‐105/AS‐017 for the BP‐01,  
AS‐108/AS‐115 for the BP‐02, AS‐017 for the FP‐01 and AS‐108/ 
AS‐115 for the FP‐02; Viton, EPDM or Kalrez.  
All wetted parts have to be absolutely clean, don’t use 
oil or grease and avoid dust! Replace the Blind/Flow plate 
and tighten the bolts by turning clockwise, tighten crosswise with 
M3 0,9 Nm for the Blind/Flow plate body valve/filter. For the 
Blind/Flow plate controller/meter with M4 2,0 Nm. Now the  
Blind/Flow plate has to be tested for leakage and function. 

 
Flow plate provides a flow path, in case where a module is removed 
temporarily or continuously from a Base Body for a valve/filter or a  
Base Body for a meter/controller.  
 
Blind plate blocks the flow, in case where a module is removed  
temporarily or continuously from a Base Body for a valve/filter or a 
Base Body for a meter/controller. In the Blind plate version for the  
valve/filter there is an O‐ring AS‐105 which can be removed when 

you want to use the plate as a Flow plate. 
In the blind controller/meter version is an expander,   

          

witch is pressed into the plate and can not be removed. 
In the flow version the expander is not present. 
 

 

Available Flow/Blind plates

 