Blind/Flow plates
-
Removal
Ensure that the system is depressurised before dismounting
the blind/flow plates. Remove the Blind/Flow plates assy by
loosening the 4 bolts counter clockwise to release the assy.
9.17.057
page 16
FP‐XX
BP‐XX
-
Replacement
Replace the gaskets AS‐105/AS‐017 for the BP‐01,
AS‐108/AS‐115 for the BP‐02, AS‐017 for the FP‐01 and AS‐108/
AS‐115 for the FP‐02; Viton, EPDM or Kalrez.
All wetted parts have to be absolutely clean, don’t use
oil or grease and avoid dust! Replace the Blind/Flow plate
and tighten the bolts by turning clockwise, tighten crosswise with
M3 0,9 Nm for the Blind/Flow plate body valve/filter. For the
Blind/Flow plate controller/meter with M4 2,0 Nm. Now the
Blind/Flow plate has to be tested for leakage and function.
A Flow plate provides a flow path, in case where a module is removed
temporarily or continuously from a Base Body for a valve/filter or a
Base Body for a meter/controller.
A Blind plate blocks the flow, in case where a module is removed
temporarily or continuously from a Base Body for a valve/filter or a
Base Body for a meter/controller. In the Blind plate version for the
valve/filter there is an O‐ring AS‐105 which can be removed when
you want to use the plate as a Flow plate.
In the blind controller/meter version is an expander,
witch is pressed into the plate and can not be removed.
In the flow version the expander is not present.