beautypg.com

4 advanced operation, 1 stored calibration curves, 2 multi fluid / multi range instruments – Bronkhorst FLOW-SMS Series User Manual

Page 21: 4 sealing material

background image

BRONKHORST HIGH-TECH B.V.

9.17.057

page 21

4 Advanced operation

4.1 Using mass flow meters/controllers on different gases or under different
conditions

4.1.1 Stored calibration curves

FLOW‐SMS instruments with digital pc‐boards allow storage of max. 8 calibration curves for various gases or process 
conditions. Fluidsets can digitally be selected when the instrument is hooked‐up by RS232 to a computer, using 
(Bronkhorst) software tools. 

4.1.2 Multi Fluid / Multi Range instruments

Instruments with activated Multi Fluid / Multi Range (MFMR) functionality are calibrated for standard ranges which 
can easily be configured for other fluids and ranges. Changing fluid and range can be performed by means of the 
software tool FlowTune, through the RS232 connection of an instrument. The program can convert the primal 
calibration curve inside the instrument to the selected fluid and range. 
In digital instruments with activated MFMR functionality the parameters capacity, density, unit type, capacity unit etc. 

are dynamic parameters. For example, changing the capacity unit from ‘ml

n

/min’ to ‘l

n

/min’ effects that the capacity 

automatically changes, for instance from 2000 ml

n

/min to 2 l

n

/min. The 100% output is not affected. Changing the 

capacity from 2000 ml

n

/min to 1000 ml

n

/min effects that the instruments full scale capacity (100% output) changes 

to 1000 ml

n

/min. The instrument is reranged. 

4.1.3 Conversion factor calculation using Fluidat software

Bronkhorst High‐Tech B.V. gathered the physical properties of over 600 fluids in a database called FLUIDAT.  
Application software, such as FLOW CALCULATIONS, enables the user to calculate accurate conversion factors, not 
only at 20°C/1 atm but at any temperature/pressure combination, both for gases and for liquids. 
Apply to your distributor for more details of this software.  
 

 

Check FLUIDAT®  on http://www.fluidat.com 
FLUIDAT® is a collection of routines to calculate physical properties of gases and liquids. These routines 
are made available at the FLUIDAT® on the Net website. 

4.1.4 Sealing material

 

Before using for a different fluid, check if the sealing material of the complete module like O‐rings, 
plunger and packing gland of capillary are suitable for the application. The instrument is fitted with 
Viton, EPDM or Kalrez seals. Which sealing material is used is shown in the last character of the model 
key. It is also important to make sure that the seals of the mounting accessories (base block, adapters, 

etc.) are suitable for the fluids used (see schematic diagram /dimensional drawing).