3 troubleshooting summary general – Bronkhorst FLOW-SMS Series User Manual

Page 24

background image

BRONKHORST HIGH-TECH B.V.

9.17.057

page 24

5.3 Troubleshooting summary general


Symptom 

Possible cause 

Action 

 

 

 

No output signal 

No power supply 

Check power supply and hook‐up 

 

 

Check cable connection and hook‐up. 

 

 

Check status of LED’s 
(see manual 9.17.023 chapter 10) 

 

No setpoint  

For analog used instruments measure 
between pin 3 and 8 of sub‐D9 
connector. 
For digital use make sure to perform a 
read setpoint after sending 

 

Cable damaged or hooked‐up wrong 

Check and compare signals at both ends 
of cable. 

 

PC‐board damaged due to long lasting 
shortage and/or high‐voltage peaks 

Return to factory 

 

No or too low inlet pressure 

Increase inlet pressure 

 

 

Open shut‐off at inlet and outlet 

 

Supply pressure too high, or differential 
pressure across meter too high 

Reduce supply pressure 

 

Valve blocked / contaminated 

Set control mode to valve fully open 
and check if it opens 

 

Sensor / capillary failure 

Return to factory 

 

 

 

Maximum output signal 

Leakage of control valve 

Set control mode to valve close and 
check if it closes 

 

Sensor / capillary failure 

Return to factory 

 

 

 

Output signal much lower than 
setpoint signal or desired flow 

Incorrect type of gas or too low inlet 
pressure 

Test instrument on conditions for which 
it was designed 

 

 

 

Oscillation 

Supply pressure / diff. pressure too high  Reduce pressure 

 

Pressure regulator of supply pressure is 
oscillating or wrong sized 

Replace pressure regulator 

 

Controller adjustment wrong 

Adjust controller with PID settings 

 

 

 

Flow is gradually decreasing 

Condensation, can occur e.g. with NH

3

  

Decrease supply pressure and / or heat 
gas to be measured 

 

 

 

Small flow at zero setpoint 

Valve leaks due to damage or dirt 

Return to factory 

 

Increased zero reading without flow 
caused by raised zero‐point 

Perform an auto‐zero action 

 

 

 

No digital communication 

Occupied or wrong bus address 

Change address with software or 
rotary‐switches 

 

No or wrong bus termination 

Check cabling and termination