BRONKHORST HIGH-TECH B.V.

2 Starting up

2.1 Check properties

9.17.057

page 6

Before installing a FLOW‐SMS system it is important to 
read the attached labels and check
‐ Flow rates / Pressure rate(s) 
‐ Fluids to be measured 
‐ Up‐ and downstream pressures 
‐ Input/output signal 
‐ Temperature 
 
 

input/output 
 A – 0…5 Vdc 
 B – 0…10 Vdc
 F – 0…20 mA
 G – 4…20 mA 

Check if the sealing material like O‐rings, plunger and packing gland of capillary are suitable for the 
used gas and process. The Flow module is fitted with Viton, EPDM or Kalrez seals. Which sealing 
material is used is shown in the last character of the model key. The rest of FLOW‐SMS system also 
contains sealings, which are selected based on the fluid flow schematic diagram. 


2.2 Rated pressure test inspection

Each FLOW‐SMS system is pressure tested to at least 1.5 times 
the working pressure of the process conditions stipulated by the customer,  
with a minimum of 8 bar. 

Pressure Testing Sticker 

Each instrument is helium leak tested to at least 210‐9 mbar l/s Helium 
outboard.
 

2.3 Check piping

For reliable measurement always make sure the fluid stream is clean.   
Use filters to assure a moisture‐, oil‐ and particle‐free gas stream. 
Recommended pore‐size: 7 µm.  
If back flow can occur, a downstream filter and a check valve are 
recommended too. 
For particle filtering the FLOW‐SMS program offers sintered filter modules 
with porosities 0.5, 2, 7, and 15 micron 

The tested pressure is stated on the instrument with a RED COLOURED sticker. Before installation, 
make sure that the test pressure is in accordance with normal safety factors for your application. 
If there is no Pressure Testing Sticker on the device or if the test pressure is incorrect, the instrument 
should not be mounted in the process line and be returned to the factory.