beautypg.com

Solenoid operated shut off valves nc and no, 6 toggle valves, 5 solenoid operated shut off valves nc and no – Bronkhorst FLOW-SMS Series User Manual

Page 13

background image

BRONKHORST HIGH-TECH B.V.

2.13.5 Solenoid Operated shut off valves NC and NO

-

Removal 
Ensure that the system is depressurised before taking shut off 
valve apart. 
The power must be off when removing the electrical connectors. 
Remove the coil by releasing the clip on top of the coil for the 
normally closed, and the nut for the normally opened. 
Loosen the 2 bolts with hexagon head counter clockwise to 
release the valve. 

 
-

Replacement 
Replace the gaskets AS‐007 and AS‐017, Viton, EPDM 
or Kalrez. 
All wetted parts have to be absolutely clean, don’t use oil or 
grease and avoid dust! 
Replace the shut off valve and tighten the bolt with hexagon 
head by turning clockwise, tighten crosswise with M3 0,9 Nm. 
Replace the coil and replace the clip on top of the coil normally 
closed, and the nut for the normally opened. 
Replace the electrical connector. 

Now the shut off valve has to be tested for leakage and function. 

Solenoid
valve coil

VDR

LED

Connector

1

2

Available electric shut off valves 

Model number 

Description 

EV‐02‐NC‐V(iton) 

Kv=0,065 

Max deltaP 3 bar 

EV‐02‐NC‐E(PDM) 

Kv=0,065 

Max deltaP 3 bar 

EV‐02‐NC‐K(alrez) 

Kv=0,065 

Max deltaP 3 bar 

EV‐02‐NO‐V(iton) 

Kv=0,065 

Max deltaP 3 bar 




    

     

Hook‐ up solenoid operated valve.  
Supply voltage + 24Vdc 
Power consumption 3 Watt. 

2.13.6 Toggle valves

-

Removal 
Ensure that the system is pressure less before taking 
the toggle valve apart. 
Loosen the 4 bolts with hexagon head counter clockwise 
to release the toggle valve. 

 
-

Replacement 
Replace the gaskets AS‐007 and AS‐017, Viton, 
EPDM or Kalrez. 
All wetted parts have to be absolutely clean, don’t use 
oil or grease and avoid dust! 
Replace the toggle valve centre before tighten the bolt 
with hexagon head by turning clockwise, tighten crosswise 
with M3 0,9 Nm. 
Now the toggle valve has to be tested for leakage and function. 



Available Toggle valves 

Model number 

Description 

TV‐01‐V 

Kv=0,09 

Toggle Valve with Viton seals 

TV‐01‐E 

Kv=0,09 

Toggle Valve with EPDM seals 

TV‐01‐K 

Kv=0,09 

Toggle Valve with Kalrez seals 



9.17.057

page 13