5 Troubleshooting and service

5.1 General

For a correct analysis of the proper operation of a flow/pressure meter or controller it is recommended to remove the 
unit from the process line and check it without applying fluid supply pressure. In case the unit is dirty, this can be 
ascertained immediately by loosening the compression type couplings and, if applicable the flange on the inlet side. 
 
Energizing or de‐energizing of the instrument indicates whether there is an electronic failure. After that, fluid pressure 

is to be applied in order to check behaviour. If there should be suspicion of leakage in case of a gas unit, do not check 
for bubbles with a leak detection liquid under the cover as this may lead to a short‐circuit in the sensor or p.c.board. 

5.2 LED indication

The red LED on the instrument gives error or warning information. 
 

 

 

 

Red LED 

Time 

Indication 

Off 

Continuously 

No error 

Flash 

Variable 

Bus activity on the Modbus interface 

On 

Continuously 

Critical error message 
A serious error occurred in the instrument 
Instrument needs service before further using 

 

For more information check the instruction manuals for digital communication/interfaces at 
http://www.bronkhorst.com/en/downloads/instruction_manuals/ 
 