Checking the Stock Registration

If you Calibrate Stock, there is no need to calibrate the sensor.

1.

Load the supply for non-peel mode as shown below.

2.

Select Stock Registration from the Printer Diagnostics window.

3.

Select the sensor to calibrate (either Black Mark or On-Demand). The black mark sensor is
used with black mark supplies (supplies with rectangular marks on the back side of the liner
or on the supply). The on-demand sensor is for peeled supplies.

4.

For black mark supplies longer than 4.0 inches, make sure supplies are loaded with a black
mark slightly below the sensor as shown below. This feeds the least amount of labels during
the calibration process.

B l a ck m ar k on s u p p l y s l i gh t l y
b e l ow se n s or w it h e n o u g h l a b el
t o f e e d t hr o ug h p r i n t er .

B l a ck M ar k Se n s o r

5.

Close the supply cover, then touch Calibrate Sensor to calibrate the black mark sensor.

The printer feeds supply until it detects a black mark to calibrate the sensor. When the

calibration is successful, a message appears: “Black Mark stock sensor calibration

successful.” Touch OK.

Using Printer Diagnostics 6-5