Output Protection:
Output will withstand forward voltages of up to 20V above rated output
voltage. Reverse protection by diode clamp for currents up to 3A.
Over-voltage Protection:
(OVP)
Range 1V to 40V (XDL 35-5), 1V to 60V (XDL 56-4)
Resolution 0·1V; accuracy ± (2% + 0·5V)
Response time typically 100
µs
Over-current Protection:
(OCP)
Range 0·01A to 5·5A (XDL 35-5), 0·01A to 4·4A (XDL 56-4)
Resolution 0·01A; accuracy ± (0·2% + 0·01A)
Response time typically 35ms
Protection Functions:
Output trips off for OVP, OCP, over-temperature and Sense miswiring
METER SPECIFICATIONS (Main Outputs)
Display Type:
5-digit (Volts), 4-digit (Amps), 14mm (0·56") LED.
Voltage (CI mode):
Resolution 10mV
Accuracy ± (0·1% of reading + 10mV)
Current (CV mode):
Resolution 0·001A; 0·1mA on 500mA range
Accuracy ± (0·2% + 0·005A); ± (0·2% + 0·5mA) on 500mA range
V x A:
Resolution 0·01W; 0·001W on 500mA range
Accuracy ± (0·3% + 0·05W); ± (0·3% + 0·005W) on 500mA range
AUXILIARY OUTPUT (T models only)
Voltage Range:
1V to 6V
Voltage Setting:
Resolution: 10mV
Accuracy: ± 0.5% ±10mV
Current Limit:
3A minimum
Output Switch:
Electronic, non-isolating. Switch illuminated when Output on.
Output Terminals:
Universal 4mm safety binding posts on 19mm (0·75”) pitch. Duplicate
screwless terminals on rear panel.
Output Protection:
Output will withstand up to 16V forward voltage. Diode clamp reverse
protection for currents up to 1A. Over-current trip.
Ripple & Noise:
(20MHz bandwidth)
<2mV rms, 10mVp-p
Load & Line Regulation:
<1·0% for a 90% load change; 0·1% for a 10% line change.
Status Indication:
Current limit lamp. Current overload trip indication.
Meter Specifications:
(use SET/VIEW button)
Voltage Meter: Resolution 10mV, accuracy ± 0.5% ±10mV
Current Meter: Resolution 10mA, accuracy ± 0.5% ±10mA
Voltage Programming
Speed:
Maximum time required for output to settle within 1% of its total
excursion (for resistive load). Excludes command processing time.
1V to 6V: 10ms, no load and full load
6V to 1V: 10ms, no load and full load
KEYBOARD & ROTARY CONTROL
All functions, including the selection and set-up of the remote control interfaces, can be set from
the keyboard. The rotary jog control can be used to adjust output voltage and current settings in
a quasi-analogue mode.