beautypg.com

AMETEK XDL Series II User Manual

Page 42

background image

41

Step/Range

Settings

Adjust for:-

Load

Detail

17, Aux o/p

5V

5.00V ± 10mV

DVM

o/p V span

18, Aux o/p

5V

5.00V on display

DVM

V readback

19, Aux o/p

1V

1.00V on display

DVM

V readback

20, Aux o/p

5V

5.00V ± 0.5mV

DVM

V readback

21, Aux o/p

off

0.000 flashing 0.001

A readback zero

22, Aux o/p

5V

Display equals load setting

or ammeter

2.7A load, or

1.8

Ω in series

with ammeter

A readback

XDL 56-4, XDL 56-4P, XDL 56-4T & XDL56-4TP

Step/Range

Settings

Adjust for:-

Load

Detail

1, 56V 2A

0.01V 2A

10mV ± 0.5mV

DVM

o/p 0V

2, 56V 2A

0.01V 2A

0.010V on display

DVM

V readback zero

3, 56V 2A

35V 2A

35V ± 0.5mV

DVM

o/p V span

4, 56V 2A

35V 2A

35.000 on display

DVM

V readback span

5, 56V 2A

41V 2A

41.000 on display

DVM

V readback 13bit

6, 56V 2A

0.01V 2A

10mV ± 0.5mV

DVM

o/p 0V

7, 56V 2A

35V 2A

35V ± 0.5mV

DVM

o/p V span

8, 56V 2A

2V 0.001A

1mA ± 0.5mA

milli-ammeter

o/p A offset

9, 56V 2A

2V 0.001A

0.001 on display

milli-ammeter

A readback zero

10, 56V 500mA

2V 0.1mA

0.1mA ± 0.05mA

milli-ammeter

o/p mA offset

11, 56V 500mA

2V 0.1mA

flashing 0.0/0.1 on display

milli-ammeter

mA readback zero

change load

12, 25V 4A

2V 4A

4A ± 0.5mA

ammeter

o/p A span

13, 25V 4A

2V 4A

4.000 on display

ammeter

A readback span

14, 56V 500mA

2V 400mA

400mA ± 0.05mA

ammeter

o/p mA span

15, 56V 500mA

2V 400mA

400.0 on display

ammeter

mA readback span

16, 56V 500mA

2V 410mA

410.0 on display

ammeter

mA readback 13bit

17, Aux o/p

5V

5.00V ± 10mV

DVM

o/p V span

18, Aux o/p

5V

5.00V on display

DVM

V readback

19, Aux o/p

1V

1.00V on display

DVM

V readback

20, Aux o/p

5V

5.00V ± 0.5mV

DVM

V readback

21, Aux o/p

off

0.000 flashing 0.001

A readback zero

22, Aux o/p

5V

Display equals load setting

or ammeter

2.7A load, or

1.8

Ω in series

with ammeter

A readback