Unit size
500
700
Miscellaneous (continued)
UF V (230 V) with
optional water cool-
ing
Filling weight of refrigerant R 404a
(1
st
stage cooling, GWP 3750)
kg
--
0.270
Filling weight of refrigerant R
508B
(2
nd
stage cooling, GWP 12300)
kg
--
0.300
UF V-UL (115 V)
Filling weight of refrigerant R 404a
(1
st
stage cooling, GWP 3750)
kg
0.375
0.365
Filling weight of refrigerant
R 508B (2
nd
stage cooling,
GWP 12300)
kg
0.370
0.410
UF V-UL (208-240 V)
Filling weight of refrigerant R 404a
(1
st
stage cooling, GWP 3750)
kg
0.375
0.360
Filling weight of refrigerant
R 508B (2
nd
stage cooling,
GWP 12300)
kg
0.370
0.380
Noise level
dB (A)
49
49
Electrical Data UF V (230 V)
IP system of protection acc. to EN 60529
IP
20
20
Nominal voltage (±10%) 50 Hz
V
230 1N~
230 1N~
Nominal power
kW
1.1
1.1
Nominal current
A
4.8
4.8
Energy consumption at -80 °C/ -112 °F, with an ambient
temperature of +25 °C / 77°F; ± 10%
kWh/day
12.8
16.2
Average heat dissipation at set-point -80 °C / -112 °F
with an ambient temperature of +25 °C / 77°F 2)
W
535
680
IEC connector plug and cable
mm / inch
2000 / 78.7
2000 / 78.7
Power plug
Shock-proof plug
Installation category acc. to IEC 61010-1
II
II
Pollution degree acc. to IEC 61010-1
2
2
Over-current release category C, 2 poles
A
10
10
Electrical Data UF V-UL (115 V)
IP system of protection acc. to EN 60529
IP
20
20
Nominal voltage (±10%) 50 Hz
V
115 1N~
115 1N~
Nominal power
kW
1.1
1.1
Nominal current
A
9.6
9.6
Energy consumption at -80 °C/ -112 °F, with an ambient
temperature of +25 °C / 77°F; ± 10%
kWh/day
***
***
Power cable (fix)
mm / inch
1800 / 70.9
1800 / 70.9
Power plug
NEMA
5-20 P
5-20 P
Installation category acc. to IEC 61010-1
II
II
Pollution degree acc. to IEC 61010-1
2
2
Over-current release category C, 2 poles
A
16
16
Electrical Data UF V UL (208 - 240 V)
IP system of protection acc. to EN 60529
IP
20
20
Nominal voltage (±10%) 60 Hz
V
208 - 240 2~
208 - 240 2~
Nominal power
kW
1.5
1.5
Nominal current
A
6.5
6.5
Energy consumption at -80 °C/ -112 °F, with an ambient
temperature of +25 °C / 77°F; ± 10%
kWh/day
***
***